Chớp cháy cốc kín D93

PDFEmail
Pensky–Martens Flash Point Testing Equipment
ASTM D93, IP 34, ISO 2719

Pensky–Martens Flash Point Testing
Pensky-Martens is the most popular method as described in the standards above. It is made up of 3 parts ‘A’, ‘B’ and ‘C’. Method ‘A’ is for testing fuels and light to medium oils. Method ‘B’ is generally for samples which are solid or highly viscous at room temperature. Method ‘C’ is applicable for biodiesel. The sample is heated and stirred at a steady rate of temperature increase which is specified according to the material being tested. The ignition source is applied at regular intervals until the vapours of the sample ignite, indicating the flash point temperature.

It is used to determine the flash point of petroleum products in the temperature range from 40 to 370°C by a manual or automated Pensky-Martens closed-cup apparatus, and biodiesel in the temperature range of 60 to 190°C by an automated Pensky-Martens closed cup apparatus.
Rating: Not Rated Yet
Manufacturer: Stanhope - Seta
Description

Seta PM-93 35000-0
Automated Pensky-Martens Flash Point
ASTM D93 Procedures A, B & C; IP 34 Procedures A, B & C;
ISO 2719 Procedures A, B & C

The Seta PM-93 is a fully automated Pensky-Martens closed cup flash point tester which combines strict method conformance with the latest control technology and safety systems to provide the ultimate level of fast, accurate and safe analysis. 

Flash points can be determined up to 400C using either standard mode for where the flash point is known or search mode for when it is not. Once complete, results are graphically displayed on the LCD screen and can then be saved to the internal memory, USB, emailed or transferred via QR code.

SQC software allows analysis of results, calculating mean, standard deviation, standard error and repeatability, plotting results in accordance with ASTM D6299.

 

SPECIFICATIONS

Ignition system: Electric or gas
Flash detection: Thermal
Cooling: Forced air (integral fan)
Heating Rate: Fast heating rate (>10°C/min) and standard; 5.5°C/min; 3°C/min; 1.3°C/min; 1°C/min
Temperature range: Ambient +5°C to 400°C
Internal Memory: 30 programmable test profiles, operator names, test methods and unlimited tests results
Fire Detection: Thermal with optional integral Fire Extinguisher
Size/Weight: 38.5 x 24 x 50 cm / 25 kg
Voltage 230 Vac +/-15% or 110 Vac +/-15%. Frequency 50/60 Hz. Auto-sensing - no selector switch required
CCCN Code: Tariff 90268020